Cavitron® Plus™Ultrasonic ScalerCavitron® Plus™Ultrasonic ScalerInstallation and Service ManualPlease read carefully and completely before operating u
9Power IndicatorLights (3 Sec. delay) when the Main Power ON/OFF Switch is ON (“I” position). Low Batte
The many styles of Cavitron and Cavitron Bellissima 30K Ultrasonic Inserts are easily interch
Pressing anywhere on the top of the foot control activates the system. • 1.
For optimal access to both the upper and lower arches, the backrest of the chair should be adjusted as for other dental procedu
1. Remove ultrasonic insert used. Clean and sterilize the ultrasonic insert(s) following the procedures outlined in th
2. Check that the system’s power cord plug is fully seated in an approved AC wall outlet. 3. Check that the wall outlet is functional.Sympt
The Cavitron Plus Ultrasonic Scaler is warranted for TWO YEARS from date of purchase. The Steri-Mate Handpiece enclosed w
.PURGEIlluminates when the Main Power On/Off switch is in the “ON” posi
1. Check that the Main Power Switch is in the ON (I) position, and that the detachable power cord is fully seated in the receptacle o
™This troubleshooting section is meant for use by qualified Cavitron® Service Technicians.
™This troubleshooting section is meant for use by qualified Cavitron® Service Technicians.
™This troubleshooting section is meant for use by qualified Cavitron® Service Technicians.
™This troubleshooting section is meant for use by qualified Cavitron® Service Technicians.
™This troubleshooting section is meant for use by qualified Cavitron® Service Technicians.
™This troubleshooting section is meant for use by qualified Cavitron® Service Technicians.
CAUTION: THIS UNIT CONTAINS COMPONENTS WHICH ARE SUBJECT TO ELECTROSTATIC DISCHARGE (ESD) DAMAGE. THE
8. Lift the protective cover from the Power Entry module. The fuses are mounted vertically under the cover. Use a small flat blade scr
10. Position the replacement Power Supply assembly with the RED-BLK wire connector at the rear. 11. Slide the Power Supply assembly under th
1. Follow the steps above in Section A for “Top Cover Removal”. 2. Unplug the Top Cover ribbon cable at
3 3 3
33
Manufactured by:DENTSPLY ProfessionalDENTSPLY International1301 Smile WayYork, PA 17404-1785Distributed by:DENTSPLY CanadaWoodbridge, OntarioL4L 4A3U
Congratulations!Your decision to add the Cavitron® Plus™ Ultrasonic Scaler to your practice represents a wise investment in good dentistr
All general supra and subgingival scaling applicationsPeriodontal debridement for all types of periodontal diseasesEn
Like bristles of a toothbrush, ultrasonic insert tips “wear” with use. Inserts with just 2 mm of wear lose about 50% of their scaling efficienc
Carefully unpack your Cavitron Plus System and verify that all components and accessories are included: 1. Cavitron® Plus™
To remove the water line from the Cavitron Plus System, turn off the dental office water supply. Disconnect the water supply line from the dent
Main Power ON/OFF Switch ON/OFF Switch located at the center front underside of the system. Ultrasonic Power Level ControlTurn knob to select the ult